Leica SP8 STED

Manufacturer:
Leica

Model name:
SP8 STED/FLIM/FCS  

Location:
HPM C35

Phone:
3 75 88

external pageDirect booking link to PPMS
 

 

 

Instrument contact:

Joachim Hehl
  • HPM C 52.1
  • +41 44 633 62 02

ScopeM
Otto-Stern-Weg 3
8093 Zürich
Switzerland

Enlarged view: SP8_STED

Brief description:

This confocal Leica TCS SP8 microscope consists of  an inverted stand equipped with a STED 3X unit (STimulated Emission Depletion) and a setup for time resolved fluorescence microscopy including a FLIM (Fluorescnece LIfetime Microscopy) and a  FCS/FCCS unit (Fluorescence /Cross/ Correlation Spectroscopy).

Microscope:

Leica DMI6000B-CS with blue/green/red epifluorescence illumination for visual observation through the eyepiece; Condensor (Air 0.55NA)

Illumination:

  • Laser unit for confocal acquisition (AOBS system): 458, 477, 488, 496, 514 nm (Argon laser), 405 nm, 561 nm, 633 nm
  • White Light Laser: tunable 470–670 nm
  • STED Lasers: 592 nm, 660 nm, 775 nm
  • FLIM Laser: 440 nm

Objectives:

  • 10x 0.4NA HC PLAN APO CS
  • 20x 0.75 NA, Oil/W/Glyc, HC PLAN APO CS2
  • 63x 1.4NA Oil  HC PL APO CS2
  • 63x 1.2NA W HC PL APO CS2
  • 100x 1.4NA Oil  HC PL APO CS2

Detectors:

  • 1 HyD
  • 2 HyD SMD (Single Molecule Detection)
  • 2 PMT
  • 1 PMT Trans

FLIM/FCS/FCCS unit from Picoquant:

  • PicoHarp 300 (TCSPC Module (Time Correlated Single Photon Counting system))
  • Pulsed Diode Laser (PDL 800-B) 440 nm
  • 4 Channel Detector Router (PHR 800)
  • Software: SymPho Time 64 1.6

Software:

  • Leica LAS X SP8 Version 1.0 (offline Leica LAS AF version 3.3 at Image processing station 2 in HPI F10.1)
  • additional licences for STED, Matrix (with CAM) for HCS applications, Life data mode, Dye separation, 3D rendering
  • Huygens Deconvolution
  • FLIM/FCCS Software: SymPho Time 64 1.6 (Picoquant)

Key features:

  • Stage: Motorized Märzhäusser stage with Leica Galvo stage
  • Scanner: Tandem Scanner with Galvanometric scanning unit and Resonant scanning unit (8 kHz)
  • Incubation: LIS box type temperatur incubation (Cube) with CO2 (Fitting Ibidi, LabTeks and 35 mm dishes, Mikrotiteplates)

Fluorescence settings:

  • For internal detection: Leica patented flexible filterfree detection within whole VIS range, spectral detection possible
  • Fluorescence filter cubes for DAPI, FITC, TRITC for visual inspection